Previous page Next page

Verification of trueness and precision of the analytical method using the IAEA approach

As part of the internal quality control of the analytical method, the measurement of a Certified Reference Material (CRM) serves to investigate if the obtained measurement result is or is not significantly biased in relation with the reported reference value. For evaluating the bias, a validated trueness and precision test (the IAEA approach) can be applied.

For trueness evaluation, the results are considered "Acceptable" if:

where:


Vref - reference value of CRM

Vlab - value obtained in laboratory

U2ref,lab - standard combined uncertainties associated to Vref and Vlab

For evaluation of precision, the results are considered "Acceptable" if:

where:

ALP: Acceptable Limit of Precision, set for each analyte as a function of its concentration level and ease of determination.

A result must obtain an "Acceptable" score in both criteria to be assigned the final score of "Acceptable".

See an excel example here.

Please check Microsoft Excel macro's security settings in order to allow code execution before opening the provided document.
previous page next page